Size | 9.06 x 3.15 x 2.36 in (230 x 80 x 60 mm) |
---|
|
Weight | 24.7 oz (700 g) |
---|
|
Sensor | Photo-ionization sensor with standard 10.6 eV lamp
(9.8 eV lamp in MP186)* |
---|
|
Response Time | 3 seconds (t90) |
---|
|
Accuracy | ±3% (at calibration point) |
---|
|
Battery/ Run Time | Rechargeable Lithium-Ion battery with 24 hours typical
operation |
---|
|
Keypad | 4 Operation keys |
---|
|
Sampling Pump | Built-in pump with 3 settings from 300 to 430 cc/min
Sample from up to 100 ft (30 m) |
---|
|
Display | 128 x 128 graphical LCD, 1.77 x 1.73 in (45 x 44 mm),
with LED backlight for enhanced display readability |
---|
|
Direct Readout | Real-time reading of gas concentration (ppb, ppm,
mg/m3, μg/m3), PID measurement gas and correction
factor, lamp on/off, Man-Down alarm on/ off, battery
status, pump status, datalogging on/off, wireless
on/off, temperature and time |
---|
|
Alarms | Audible (90 dB @ 30 cm), visual (flashing bright red
LEDs), and on-screen indication of alarm conditions
plus wireless remote alarm notification
• High, Low, TWA and STEL alarms
• Over range alarm, battery low alarm
• Man-Down alarm with pre-alarm and real-time
remote wireless notification |
---|
|
EMI/RFI | Highly resistant to EMI/RFI
Compliant with EMC Directive 89/336/EEC |
---|
|
IP Rating | IP-66/67 |
---|
|
Datalogging Capacity | Standard 12 months at one-minute intervals
Storage interval adjustable from 1 to 3,600 seconds |
---|
|
Calibration | Two/three-point calibration |
---|
|
Low Flow Alarm | Auto pump shutoff at low-flow condition |
---|
|
Charging and Communication | Charging, data download, instrument setup and firmware
upgrades on PC or laptop via Micro USB
Wireless data and alarm status transmission via builtin RF modem |
---|
|
Wireless Range | 1,000 ft (300 m) line of sight |
---|
|
Correction Factors | Integrated Correction Factor list of more than 200
compounds |
---|
|
Safety Certifications | Class I, Div 1, Group ABCD
Class II, Div 1, Group EFG
Class III, Div 1
T4, -20°C ≤ Tamb ≤ +50°C
Ex ia IIC T4 Ga
II 1G Ex ia IIC T4 Ga (pending)
European Conformity
UL Listed |
---|
|
Temperature | -4° to 113°F (-20° to 50°C) |
---|
|
Humidity | 0% to 95% Relative humidity (non-condensing) |
---|
|
Attachments | Durable rubber boot, color coded for different models;
Tube holder for MP186 |
---|
|
Warrenty | 2 Years including lamp and sensor